CalRMethod
The receiver characterization method the PNA should use during a noise calibration.
The valid possible values are:
- PowerMeter
- NoiseSource
The default value is: NoiseSource
CalRMethod is used in the following calibration:
If not specifically set to 'PowerMeter' or it is omitted the default is NoiseSource
This key corresponds with the PNA SCPI command "SENSe<ch>:NOISe:CALibration:RMEThod <string>" parameter as explained in the Keysight Documentation.
An Example:
;*******************************************************************************
; Example configuration for a calibration type Noise using 2 ports
;*******************************************************************************
[CAL_Noise_2Port]
CalAcquireTimeoutSec = 200
CalRMethod = NoiseSource
CalSetDescription = CalSet for CAL_Noise_2Port
CalType = Noise
ConfigType = Calibration
CsaLoadFilename = Before_CAL_Noise_2Port.csa
DirectoryToUse = C:\Examples
NoisePath = PATH_NoiseSource
PathName = PATH_Noise_2Port
SaveFilename = After_CAL_Noise_2Port.csa
ThruDelay1To2 = 3E-09
ThruPath1To2 = PATH_UnknownThru_1-2
ThruPortPairs = 1:2
; FCM calibration device definitions
[ICM_FCM1]
ConfigType = Fcm
IpAddress = 10.11.0.201
[ICM_FCM2]
ConfigType = Fcm
IpAddress = 10.11.0.202
; ICM MultiSwitch controller and switch definitions
[ICM]
ConfigType = MultiSwitch
IpAddress = 10.11.0.100
[ICM_SW01]
ConfigType = Switch
Device = ICM
SwitchIndex = 1
[ICM_SW02]
ConfigType = Switch
Device = ICM
SwitchIndex = 2
[ICM_SW03]
ConfigType = Switch
Device = ICM
SwitchIndex = 3
[ICM_SW04]
ConfigType = Switch
Device = ICM
SwitchIndex = 4
[ICM_SW05]
ConfigType = Switch
Device = ICM
SwitchIndex = 5
[ICM_SW06]
ConfigType = Switch
Device = ICM
SwitchIndex = 6
[ICM_SW07]
ConfigType = Switch
Device = ICM
SwitchIndex = 7
[ICM_SW08]
ConfigType = Switch
Device = ICM
SwitchIndex = 8
[ICM_SW09]
ConfigType = Switch
Device = ICM
SwitchIndex = 9
[ICM_SW10]
ConfigType = Switch
Device = ICM
SwitchIndex = 10
; Path defining FCMs to be used during calibration
[PATH_Noise_2Port]
ConfigType = Path
PnaPort1Device = ICM_FCM1:1
PnaPort2Device = ICM_FCM2:1
SwitchPath = PATH_TestPortThru
; Path to the noise source
[PATH_NoiseSource]
ConfigType = Path
SwitchPath = ICM_SW10:1
; Path for test ports through to FCMs
[PATH_TestPortThru]
ConfigType = Path
SwitchPath = ICM_SW01:1,ICM_SW02:1,ICM_SW03:1,ICM_SW04:1,ICM_SW10:0
; Path for unknown thru between port 1 and 2
[PATH_UnknownThru_1-2]
ConfigType = Path
SwitchPath = ICM_SW01:2,ICM_SW02:2,ICM_SW03:2,ICM_SW04:2,ICM_SW05:4,ICM_SW06:1,ICM_SW07:1,ICM_SW08:4
; PNA definition
[PNA-x]
ConfigType = Pna
DirectoryToUse = C:\Temp
IpAddress = 10.11.0.200