This is the description that will be given to the PNA Cal Set associated with this calibration. Having a good description helps identify the Cal Set when looking it up on the PNA.

There is no default value, but it is not required.

All Calibration types use this value

This key can be left blank or omitted, however this is not recommended.

This key corresponds with the PNA SCPI command "SENSe<cnum>:CORRection:CSET:DESCription <string>" parameter as explained in the Keysight Documentation

An Example:

;*******************************************************************************
; Example configuration for a calibration type SParam using 1 port
;*******************************************************************************
[CAL_SParam_1Port]
CalAcquireTimeoutSec = 60
CalSetDescription = Cal Set for CAL_SParam_1Port calibration
CalSetName = CAL_SParam_1Port
CalType = SParam
ConfigType = Calibration
DirectoryToUse = C:\Examples
LoadFilename = Before_CAL_SParam_1Port.csa
PathName = PATH_SParam_1Port
PnaDisplayEnable = 1
SaveFilename = After_CAL_SParam_1Port.csa

; FCM calibration device definition
[FCM_1]
ConfigType = Fcm
ConnectionTimeoutSec = 1.0
FileTransferTimeoutSec = 60
IpAddress = 10.11.0.201
ResponseTimeoutSec = 60

; Path referenced by calibration defining which device will be used
[PATH_SParam_1Port]
ConfigType = Path
PnaPort1Device = FCM_1:1

; PNA definition
[PNA-X]
ConfigType = Pna
ConnectionTimeoutSec = 1.0
DirectoryToUse = C:\Temp
FileTransferTimeoutSec = 60
IpAddress = 10.11.0.200
PortCount = 4
ResponseTimeoutSec = 60
VerifyCommands = 0