The PNA directory where files will be located that are transfered to/from the PNA.

It is recommended that this key be set otherwise files are saved to a default directory which may be difficult to find.

When a calibration is run the files from the calibration devices used are cached to the PNA so that they can be accessed quickly. These files are cached in the directory specified.

If a configuration key defines a full path, this value is not used. If a configuration key does not define a full path, but just a file name, this value is used.

An Example:

;*******************************************************************************
; Example configuration for a calibration type SParam using 1 port
;*******************************************************************************
[CAL_SParam_1Port]
CalAcquireTimeoutSec = 60
CalSetDescription = Cal Set for CAL_SParam_1Port calibration
CalSetName = CAL_SParam_1Port
CalType = SParam
ConfigType = Calibration
DirectoryToUse = C:\Examples
LoadFilename = Before_CAL_SParam_1Port.csa
PathName = PATH_SParam_1Port
PnaDisplayEnable = 1
SaveFilename = After_CAL_SParam_1Port.csa

; FCM calibration device definition
[FCM_1]
ConfigType = Fcm
ConnectionTimeoutSec = 1.0
FileTransferTimeoutSec = 60
IpAddress = 10.11.0.201
ResponseTimeoutSec = 60

; Path referenced by calibration defining which device will be used
[PATH_SParam_1Port]
ConfigType = Path
PnaPort1Device = FCM_1:1

; PNA definition
[PNA-X]
ConfigType = Pna
ConnectionTimeoutSec = 1.0
DirectoryToUse = C:\Temp
FileTransferTimeoutSec = 60
IpAddress = 10.11.0.200
PortCount = 4
ResponseTimeoutSec = 60
VerifyCommands = 0