The name of the Cal Set that will be used on the PNA to store the error terms of this calibration.

The default value is the calibration name.

All Calibration types should set this if a value other than the default is desired. This key is not required and can be omitted, however, if it is not present or is blank, the default name of the Cal Set will be the same as the calibration name.

It is good practice to name each cal set the same as the calibration names so that they can be looked up on the PNA more easily.

This key corresponds with the PNA SCPI command "SENSe<cnum>:CORRection:CSET:NAME <string>" parameter as explained in the Keysight Documentation

An Example:

;*******************************************************************************
; Example configuration for a calibration type SParam using 1 port
;*******************************************************************************
[CAL_SParam_1Port]
CalAcquireTimeoutSec = 60
CalSetDescription = Cal Set for CAL_SParam_1Port calibration
CalSetName = CAL_SParam_1Port
CalType = SParam
ConfigType = Calibration
DirectoryToUse = C:\Examples
LoadFilename = Before_CAL_SParam_1Port.csa
PathName = PATH_SParam_1Port
PnaDisplayEnable = 1
SaveFilename = After_CAL_SParam_1Port.csa

; FCM calibration device definition
[FCM_1]
ConfigType = Fcm
ConnectionTimeoutSec = 1.0
FileTransferTimeoutSec = 60
IpAddress = 10.11.0.201
ResponseTimeoutSec = 60

; Path referenced by calibration defining which device will be used
[PATH_SParam_1Port]
ConfigType = Path
PnaPort1Device = FCM_1:1

; PNA definition
[PNA-X]
ConfigType = Pna
ConnectionTimeoutSec = 1.0
DirectoryToUse = C:\Temp
FileTransferTimeoutSec = 60
IpAddress = 10.11.0.200
PortCount = 4
ResponseTimeoutSec = 60
VerifyCommands = 0